-50...1600°C IR spindulių ir kontaktinis termometras DIT-500 su 50:1 optine skyra, K tipo termopora, duomenų registravimu ir jungimu su PC

Gamintojas: Sonel
Prekės kodas: WMGBDIT500
Kaina:
347,00 €be PVM

Temperature meters operating in the infrared are used to determine the temperature of the surface of the item. The optical system of the device detects the emitted, bounced and sent energy, which is collected and concentrated on the detector. Electronic system translates optical data into temperature value. In order to increase accuracy the laser pointer is built-in the device.

 

IR temperature range

IR temperature
range
D:S Resolution Accuracy
-50...999,9°C
-58...999,9°F
50:1 0,1°C
0,1°F
-50...20°C
-58...68°F
±2,5°C
±4,5°F
1000...1600°C
1000...2912°F
1°C
1°F
20...400°C
68...752°F
±(1,0% m.v. + 1°C)
±(1,0% m.v. + 1,8°F)
400...800°C
752...1472°F
±(1,5% m.v. + 2°C)
±(1,5% m.v. + 3,6°F)
800...1600°C
1472...2912°F
±2,5% m.v.

 

Type K temperature range

Type K
temperature range
Resolution Accuracy
-50...999,9°C
-58...999,9°F
0,1°C
0,1°F
±(1,5% m.v. + 3°C)
±(1,5% m.v. + 5°F)
1000...1370°C
1000...2498°F
1°C
1°F
±(1,5% m.v. + 2°C)
±(1,5% m.v. + 3,6°F)
D:S” - Distance (D) to Spot Size (S)
„m.v.”- measured value
 
Rated operational conditions:
- operating temperature: 0...+50°C
- storage temperature: -20...+60°C
- humidity: 10...90%

Other technical data:
- display LCD: segment with a backlight
- spectral response: 8~14μm
- emissivity: digitally adjustable from 0,10 to 1,0
- polarity: automatic, (-) sign for negative polarity
- diode laser: output <1mW, wave length 630~670nm, class 2 laser product
- power supply: 9V battery, NEDA 1604A or IEC 6LR61
 
„CE” - comply with EMC
 

- over range indication: LCD will show „----”
- response time: 150 ms
- weight: 350 g
- dimensions: 230 x 155 x 54 mm